CERTIFICATE OF CALIBRATION
FOR BRÜEL & KJÆR SOUND INTENSITY MICROPHONE


The calibration is performed by comparison with                                 Microphone Cartridge Type                                  4181

Calibration Service Standard Microphone Cartridge Type                    Part  #  1        Id No. N/A                   Serial No.    1480271

 4160   Serial No.  991820                                                                      Open Circuit Sensitivity

Calibrated by   TS (BRÜEL & KJÆR)         Date  14 AUG 2000                        at 1013 mbar, 23 o C, 65% R.H.:   -38.98  dB re. 1 V/Pa

                                                         Due  14 AUG 2001                                                              or               11.25       mV/Pa

a)  Estimated uncertainty of comparison:                                              This calibration is traceable to:

      + 0.09  dB at 99% confidence level.                                                 NIST Test # 822/261669-99

b)  Estimated uncertainty of Calibration Service                                                   

      Standard Microphone Cartridge:                                                      Open Circuit Correction Factor K0 =  +13.0 dB

      + 0.06  dB at 99% confidence level.                                                 Correction Factor K =          dB

c)  Absolute uncertainty:                                                                        with Preamplifier Type:

       Sq. Root (a2+b2) = 0.11  dB at 99% confidence level.                     Serial No.





The lower curve is the pressure response recorded                              Condition of Test:

with Electrostatic Actuator Type :  UA 0033                                        Ambient Pressure:                  988.71                     hPa

The Free-Field Frequency Response Curve (0o) (if                         Temperature:                          23                            oC

shown) is obtained by applying the corrections, found                         Relative Humidity:                 38                            %

on the supplied chart, to the measured actuator                                     Polarization voltage:               200                          V

response curve.                                                                                       Frequency:                              250                          Hz

The Random Incidence Frequency Response Curve                        Date of calibration:                 08 MAY 2001

(if shown) is obtained by applying the corrections,                               Re-calibration due:                  08 MAY 2002

found on the supplied chart to the measured actuator                            Note: The calibration data is both “AS FOUND” and  “AS FINAL”

response curve.                                                                                       Certificate  # OM20010508   PO20010508


ODIN METROLOGY, INC.                  For: Odin Metrology, Inc., Thousand Oaks, CA 91320

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               

CALIBRATION OF BRÜEL & KJÆR INSTRUMENTS                                   Note:  At the time of calibration this microphone was
3533 Old Conejo Road Suite 125                                                                                                         performing  within  Mfg. specification.
Thousand Oaks, CA   91320                                                                                                                  Calibration performed by:
Tel. (805) 375-0830      Fax (805) 375-0405                                                                                     Torben Ehlert, Quality Assurance Manager

Note: This calibration report shall not be reproduced, except in full, without written consent by Odin Metrology, Inc.                                                                     Page 1 of 3

 